Ventilation modes |
- Assisted/controlled pressure control
- Assisted/controlled volume control
- CPAP
- BILEVEL
- SIMV
- MMV
- PRVC
- Apnoe Backup Ventilation
- Option NIV
|
Options |
- Optional adaptive cuff pressure control
- Optional patient-side CO2- and O2-measuring
- Optional external display
|
Ventilator settings |
| Tidal volume |
50 - 2500 mL |
| Minute volume |
0.5 - 50 L/min |
| Insp. flow |
0 - 180 L/min |
| Insp. pressure |
0 - 90 mbar |
| Breath rate |
1 - 120 /min |
| Insp. time |
0.1 - 15 s |
| I:E |
1:6 - 6:1 |
| Ramp |
0 - 10 s |
| PEEP |
0 - 50 mbar |
| FiO2 |
21 - 100 % |
| Sigh |
1/10 - 1/200 breaths, in a/c modi |
| Insp./Exp. pause |
15 s |
| Trigger sensitivity (flow) |
0.5 - 20 L/min |
| Trigger sensitivity (pressure) |
-1 - -10 mbar |
| Exp. flowtrigger |
5 - 90 % of insp. peak flow |
| Alarms |
Pressure, Vt, MV, flow, breaths, O2, CO2, cuff pressure |
Pressure |
| Display |
P/V, pressure-curve, PEAK, PEEP, plateau, MEAN, MIN |
| Alarms |
Pressure (high/low), PEEP (high/low) |
Flow |
| Display |
F/V, flow-curve, PEAK, insp., exp. |
| Alarms |
Flow (high/low) |
Volume |
| Display |
P/V, volume-curve, Vt, insp., exsp., leak, MV |
| Alarms |
Volume (high/low), VT (high/low), leak |
Fi/e, O2/CO2 |
| Display |
Values insp. and exsp. |
| Alarms |
Preset limits (high/low) |
Apnea |
| Alarms |
Preset limits (high/low) |
|
Apnea |
| Alarms |
Preset limits (high/low) |
Breath parameters |
| Display |
WOB, compliance, resistance, intrinsic PEEP, I:E, P 0.1 |
| Alarms |
Preset limits (high/low) |
Screen |
| Display |
High resolution 10.4 TFT-LCD-Color-Display (brightness manually or
automatically adjustable)
|
| Curves/Graphics |
Simultaneously visualized by up to 8 dynamic curves / loop diagrams |
| Data connections |
LAN oder RS232 |
Trend |
| Trend storage |
All monitored parameters, storage of max. and min values per time scale |
| Storage time |
Up to 96 hours |
| Display |
Chart graphic |
Technical data |
| Dimensions (H/W/D) |
351 * 317 * 189 mm |
| Weight |
12 kg |
| Voltage |
115 - 230 VAC / 50 - 60 Hz |
| Power consumption |
70 W |
| Internal batterie |
12 V / 30 min |
| External batterie |
12 V oder 24 V / via DC/DC-Converter |
| Gas supply |
O2 / Air by central gas supply, additional port for
aux. O2 bottle
|
| Gas pressure |
2.8 - 6 bar (280 - 600 kPa) |
Environmental conditions |
|
Operation |
Storage |
| Temperature |
10 - +40 °C |
-20 - +60 °C |
| Atmospheric pressure |
700 - 1060 hPa |
500 - 1060 hPa |
| Rel. air humidity |
0 - 90 % |
0 - 100 % |
|